三级中文字幕 I 99re6在线视频精品免费 I av超清在线 I 亚洲人色 I 午夜婷婷综合 I 蜜桃视频污污 I 在线麻豆 I 欧美成人秋霞久久aa片 I xxxx在线视频 I 成人综合av I 一区二区免费看视频 I 黄网站色视频免费国产 I 伊人国产女 I 一区二区三区免费在线 I 欧洲av在线 I 亚洲va在线va天堂 I 怡春院综合 I 18国产精品白浆在线观看免费 I 97se狠狠狠狠狼亚洲综合网 I 国产区一二 I 久久影院视频 I 日本色www I 久久99日韩国产精品久久99 I 非洲人与性动交ccoo I 成人自拍小视频 I 亚洲精品久久7777777国产 I 黄色国产网站 I 亚洲一区二区三区四区在线 I 999一区二区三区 I 2021av I 97啪啪 I 一级日本片 I 顶级欧美做受xxx000久久久

技術文章您現在的位置:首頁 > 技術文章 > 粘度計NDJ-1 英文說明書
粘度計NDJ-1 英文說明書
更新時間:2013-04-12   點擊次數:8704次

Operation Manual

For

NDJ-1 Rotary Viscometer

MADE  IN  CHINA

Thank you for using NDJ-1 Rotary Viscometer. NDJ-1 is easy to use, in order to install and operate accuray, and make full use of product features, we suggest you to read the manual before using.

· The company possesses the final interpretation power.

· The company reserves the power of amending technical rules without prior notice.

· The company reserves the power to amend the manual without further notice.

· May not be translated into other languages, reproduction prohibited without written permission. 

               

               Cataloge

1General……………………………………………………

2Main technical data…………………………………………

3Principle………………………………………………

4Installation……………………………………………………

5Operation procedures………………………

6Precaution………………………………………………

Packing List ……………………………………………………

1. General

NDJ-1 rotary viscometer is a new instrument used for determining the liquid viscose capacity and the absolute viscosity. NDJ-1 has been widely used to determine and measure the liquid viscosity in many applications such as grease, painting, plastic, pharmacy and adhesives. It is a precision instrument to monitor and control the stable quality of products in the manufacturing.

2. Main technical data

1) Measurement range: 10-100.000 mPa·s 

2) Rotor specification: 0, 1, 2, 3, 4, 5, five kind rotor (0# rotor could measure the viscometer from low viscometer to 0.1 mPa·s)

3) Measurement error: ±5% (Newton liquid)

4) Power supply: Voltage--220±22V, Frequency50 Hz±0.5Hz

5) Dimensions: 300 mm ×300 mm × 450 mm

6) Net weight: 1.5 Kg (the base not included)

3. Principle

1) Figure 1Synchronous motor rotates with stable rate, connect scale disk, through hairspring and rotating shaft, drives rotors to rotate. The rotors will subject to a torque moment proportional to liquid viscosity because of the liquid viscose hysteresis. The torque moment will be measured by the sensors and processed into the viscosity and shown on the display.

2) Rate can be changed by using gear system and clutch, operated by rotary knob, four optional velocity.

3) According to different equipments, different rotors (Five rotors: NO.0-NO.4) will be attached, it can be select with the velocity according to the viscosity of the liquid.

4) The fixed control gear of pointer are used for accurate reading. When the rate is so fast (30 rpm, 60 rpm) that you can not get the reading while it rotated, press pointer joystickfix the pointer in order to get the reading.

5) The rotor bracket is used for the protection of rotors and the stability of the measurement, a more reliable measuring result can be made by using the protection bracket. 

6) It can be used as a portable instrument, assigned fixed bracket and elevating system. Generally, it should be fixed when determine small amount or the temperature is fixed in the laboratory.

7) The instrument has gears and damping elevate mechanism. It is stable for instrument lifting.

8) The whole set of instrument has good quakeproof packing to ensure the precision of instrument.

Installation

1. Take out the bracket, package, stanchion and so on from the packaging.

2. Screw the column to the threaded hole behind the bracket, and put the tooth profile surface in the column facing front the bracket. Screw the hex nut tight by wrench to prevent the column running.

3. Rotational move the lifting chuck knob to check the flexibility and self-looking of lifting chuck. Found the lifting chuck too loosen and too tight phenomenon, can be used hex nut wrench to adjust the tight screw. So that it can lift from top to bottom. And it is suitable for tight to prevent the situation of automatically falling after installing the viscometer.

4. Take out viscometer form package and mount it on the lifting collettight it by handle setscrew to avoid loosen(as level as possible), link the power supply, take away the rubber band on the pointer joystick, screw the nut of protecting-cap loosen under the instrument, take away the protecting -cap.

5. Adjust the adjustment screw and put the rotor into the liquid to be measured till the level mark on the rotor reach the liquid surface.

Operation procedures

1. Prepare the liquid to be measured and put it into a glass beaker or a right angle container with the diameter not smaller than 70 mm, height not smaller than 130mm, take care of the liquid temperature.

2. Mount the protection bracket on instrument, turning right for mounting, turning left for removing.

3. Screw the selected rotor into connecting rod, turning right for mounting, turning left for removing , Adjust the lifting screw and put the rotor into the liquid to be measured till the level mark on the rotor reach the liquid surface. Connect power supply, turn on the equipment, screw the speed knob, select velocity, relax pointer joystick, reading data can be displayed (about 20~30 seconds) when the pointer becomes stable. Press the pointer joystick after pointer becoming stable, cut power supply, then get the reading data. If the pointer isnt in reading window after motor stop, you can press pointer joystick unceasingly, open or close the motor time after time, after practice, you can operate it familiarly.

4. When the reading data is too high or too low, the rotor or the rate can be changed, make sure the reading is between 20 and 80.

5. Choice of range, coefficient, rotor and rate:

5.1 Estimate approximay the viscose range, then select the rotor and the velocity based on the range table given below:

For example, the viscosity of liquid is about 3000 mPa·s the settings should be as following:  

rotor 2 # with velocity of 6 rpm,

rotor 3 # with velocity 0f 30 rpm

5.2 If the viscosity of liquid can not be estimated, it should take a high value for real measurement, the rotor selection should be made from smaller to larger with their number from higher to lower. In general, high viscosity should use smaller rotor with slower velocity, lower viscosity use lager rotor with faster velocity.

5.3 Coefficient: the reading must be multiplied the specific coefficient in the coefficient table in order to get the absolute viscosity.

   That is: 

            absolute viscosity

            coefficient

            reading ( deflection angle)

   5.4 Amendment of frequency error: when the power frequency is inaccurate, you can amendment it according to the formula below:

      Actual viscosity= instruction viscosity*nominal frequency/ actual frequency

Velocity

Range

Rotor

60

30

12

6

1

100

200

500

1000

2

500

1000

2500

5000

3

2000

4000

10000

20000

4

10000

20000

50000

100000

   5.5 

5.6 Coefficient table: 

        Rpm

Rotor

60

30

12

6

1

1

2

5

10

2

5

10

25

50

3

20

40

100

200

4

100

200

500

1000

 

Precaution

1. This instrument is limited to the room temperatures

2. This instrument should be used under the designed voltage and frequency and their allowable error ranges, or incorrect results could be resulted.

3. Using the bracket fix the instrument in order to determine accuray. Handheld operation: keep the instrument stable and level.

4. Care should be taken for mounting or removing rotors, slightly and lift the connecting screw bolt to avoid a transverse force acting on rotor to cause it bending.

5. The instrument mounted with rotor should not be revolved.

6. Don’t run the motor when the pointer joystick is not pressed.

7. Keep cleaned on screws and connecting points between rotor and connecting bolt rod, or a unstable rotation could be caused in the real measurement.

8. Use hand to hold the instrument when lifting up or moving down to avoid dropping it.

9. After completing measurement each time, the rotor should be fully cleaned (rotor should be removed from instrument for cleaning), then place it on the protection bracket.

10. Unauthorized removing or replacing the instrument part, and applying lubricates are not allowable.

11. When moving or shipping instrument, put on the yellow caver cap and lift the connecting bolt rod and screw the bolt on the cap tightly.

12. Suspension, emulsion or polymer and other high viscosity liquids are non-Newton liquids, their viscosity will change with shear velocity and time, it is normal for their measured inconsistent results under the selected rotor and velocity and time, and it is not resulted from the instrument problems (in general, the rotor and velocity and time should be specified for non-Newton liquids).  

13. The cautions should be taken for followings to obtain a good measuring reault:

(1) Accuray control the temperature of liquid to be measured;

(2) Make temperature of rotor and protection bracket same as liquid;

(3) Ensure the liquid homogeneity and without bubbles;

(4) Put the rotor located on the center of the liquid container in real measurement;

(5) Remove bubbles adhered on the rotor when put it into liquid;

(6) When change the rotor or rate, keep between 30~90 as possible.

(7) Use rotor protection bracket for measurement;

(8) Ensure rotor and connecting screw rod cleaned;

(9) Amend according to the amending formula when power frequency is not exact.

(10) When pointer is stable, reading with the same angle of view.

(11) Adjust level bubble exactly, move the instrument slowly, keep the instrument stable.

(12) Strictly follow the operation instruction for measurement.

Packing List

1) NDJ-1 Digital Rotary Viscometer           1 Qt.

2) Rotors, 1#, 2#, 3# and 4#                   1 Qt of each

3) Protection bracket                       1 Qt.

4) Bracket                                1 Qt.

5) Adjusting screw                         3 Qt.

6) Operation manual                        1 Qt.

7) License                                1 Qt.

8) Warranty card                           1 Qt.

上海平軒科學儀器有限公司

上海平軒科學儀器有限公司

工廠地址:上海市嘉定區南翔德園路1259弄18號

©2018 版權所有:上海平軒科學儀器有限公司  備案號:滬ICP備10019381號-1  總訪問量:1589739  站點地圖  技術支持:化工儀器網  管理登陸

主站蜘蛛池模板: 国产精品久久久久aaaa | 在线视频免费无码专区 | 久久强奷乱码老熟女网站 | 亚洲欧美日韩中字视频三区 | 免费国精产品自偷自偷免费看 | 免费观看又色又爽又黄的崩锅 | 国产成人一区二区三区影院动漫 | 亚洲国产av一区二区三区 | 国产成人无码a区视频在线观看 | av无码人妻无码男人的天堂 | 无码人妻一区二区三区在线 | 人人妻人人爽人人爽 | 中日韩中文字幕无码一本 | 国产女人爽的流水毛片 | 婷婷六月在线精品免费视频观看 | 鲁鲁狠狠狠7777一区二区 | 亚洲毛片不卡av在线播放一区 | 国产麻豆成人传媒免费观看 | 爽爽午夜影视窝窝看片 | 日产一区日产2区 | 久久中文字幕亚洲精品最新 | 精品av国产一二三四区 | 日日碰狠狠躁久久躁96avv | 中文字幕无码精品亚洲资源网久久 | 中文字幕在线观看国 | 久久国产精品人妻一区二区 | 三级黄色片免费观看 | 久久人人爽人人爽人人片av高请 | 夫妇交换性三中文字幕 | 无码内射中文字幕岛国片 | 精产国品一二三产品蜜桃 | 成人性生交大片免费看r老牛网站 | 动漫精品无码视频一区二区三区 | 国产精品亚洲二区在线观看 | a级片视频在线观看 | 成人网站免费看黄a站视频 2022久久国产精品免费热麻豆 | 欧美第一黄网免费网站 | 欧美日韩不卡高清在线看 | 热re99久久精品国产66热 | 狠狠色噜噜狠狠狠777米奇888 | 成人无码专区免费播放三区 |